15.1 Quantitative analysis and calibration methods

๐ŸŒˆSpectroscopy
Unit 15 Review

15.1 Quantitative analysis and calibration methods

Written by the Fiveable Content Team โ€ข Last updated September 2025
Written by the Fiveable Content Team โ€ข Last updated September 2025
๐ŸŒˆSpectroscopy
Unit & Topic Study Guides

Quantitative analysis in spectroscopy relies on the Beer-Lambert law, which links absorbance to concentration. Calibration curves, linear dynamic range, and detection limits are key concepts for accurate measurements. These tools help scientists determine unknown concentrations in samples.

Various calibration methods, like standard addition and internal standards, address matrix effects in complex samples. Analytical performance metrics, including precision, accuracy, and signal-to-noise ratio, ensure reliable results. These techniques are essential for chemical analysis across industries.

Quantitative Analysis Fundamentals

Beer-Lambert Law and Calibration Curves

  • Beer-Lambert law relates absorbance to concentration and path length
  • Expressed mathematically as A=ฮตbcA = ฮตbc, where A is absorbance, ฮต is molar absorptivity, b is path length, and c is concentration
  • Forms the basis for quantitative spectroscopic analysis
  • Calibration curve plots instrument response against known concentrations
  • Used to determine unknown concentrations from measured responses
  • Typically linear within a specific concentration range

Linear Dynamic Range and Detection Limits

  • Linear dynamic range spans the concentrations where the calibration curve remains linear
  • Determines the working range for quantitative analysis
  • Limit of detection (LOD) represents the lowest concentration reliably distinguished from background noise
  • Calculated as LOD=3ฯƒ/mLOD = 3ฯƒ/m, where ฯƒ is the standard deviation of the blank and m is the slope of the calibration curve
  • Limit of quantification (LOQ) indicates the lowest concentration that can be quantitatively determined with acceptable precision
  • Typically defined as LOQ=10ฯƒ/mLOQ = 10ฯƒ/m
  • LOQ is always higher than LOD, providing more reliable quantitative results

Calibration Methods

Standard Addition Method

  • Compensates for matrix effects in complex samples
  • Involves adding known amounts of analyte to the sample
  • Extrapolates the calibration curve to determine the original concentration
  • Particularly useful when matrix-matched standards are unavailable
  • Assumes a linear response and that the added standard behaves identically to the analyte in the sample

Internal Standard and Matrix Effects

  • Internal standard involves adding a known compound to both samples and standards
  • Chosen to behave similarly to the analyte but be distinguishable from it
  • Corrects for variations in sample preparation, injection, or instrument response
  • Ratio of analyte signal to internal standard signal used for quantification
  • Matrix effects occur when sample components other than the analyte influence the measurement
  • Can cause signal enhancement or suppression
  • Addressed through matrix-matched calibration or standard addition method

Analytical Performance Metrics

Precision and Accuracy

  • Precision measures the reproducibility of results
  • Expressed as relative standard deviation (RSD) or coefficient of variation (CV)
  • Calculated from repeated measurements of the same sample
  • Accuracy represents how close the measured value is to the true value
  • Often expressed as percent recovery or percent error
  • Determined by analyzing samples with known concentrations (certified reference materials)

Signal-to-Noise Ratio and Method Validation

  • Signal-to-noise ratio (S/N) compares the level of desired signal to the level of background noise
  • Higher S/N indicates better sensitivity and lower detection limits
  • Calculated as S/N=(ฮผsignalโˆ’ฮผblank)/ฯƒblankS/N = (ฮผsignal - ฮผblank) / ฯƒblank, where ฮผ represents mean values and ฯƒ is standard deviation
  • Method validation ensures analytical procedures are suitable for their intended use
  • Involves assessing various performance parameters (accuracy, precision, linearity, selectivity)
  • Crucial for regulatory compliance and quality assurance in analytical laboratories